Natural Convection & Constant Temperature Chamber

LW-9022-RT Natural Convection and Constant Temperature Chamber

LW-9022 series generate natural convection condition with accurate temperature control and good temperature uniformity. This mechanism can provide windless phenomenon to convect air simply by temperature gradients. Thus, be able to simulate authentic environments and know real heat transfer performance, including strict environmental factors. These chambers are available in different sizes. All different sizes of chambers can be applied to various kinds of thermal management and cooling issues. Programmable temperature and relative humidity (RH%) control are also available.

 

Apparatus Features

  1. Similar flow patterns with interior environment
  2. Easy Maintenance
  3. Temperature uniformity reports
 

Sample Installation

Manual Scissors Jack

  • 30 x 35 cm
  • Loading 30 kg
  • Height span 30 cm
  • For 9022 Chamber

 
 
 
 
 
 
 
 


Manual Scissors Jack with a Perforated Plate

  • 50 x 50 cm
  • Thickness 20 mm

 
 
 
 
 
 
 
 
 
 


Grid Structures

  • 2 layers
  • All grid bars are removable
  • Step-wise height adjustment

 
 
 
 
 
 
 
 
 


Suspended Platform

  • 20 x 20 cm (Reference)
  • Loading: 3 kg

 
 
 
 
 
 
 
 
 
 


Motor-Driven Scissors Jack

  • 50 x 100 cm
  • Loading 50 kg
  • Height span 23 cm
  • For 9022M

 
 
 
 
 

 

 

Example

Quality Assurance

Results are from Model: LW-9022th; RH Condition: 95%

Temperature Uniformity Report

TemperatureMeasurement PointsAverageSDNon-Uniformity
123456789°C°C
30°C29.129.429.529.229.230.229.829.329.229.450.351.18%
Deviation-0.300.1-0.2-0.20.80.4-0.1-0.2
50°C4949.84949.250.248.74949.45049.350.541.08%
Deviation-0.40.5-0.4-0.20.9-0.7-0.40.10.7
60°C59.960.960.860.560.860.860.660.660.160.540.340.56%
Deviation-0.60.40.2-0.10.20.30.10-0.4
70°C69.670.270.469.868.969.669.569.268.769.540.560.81%
Deviation00.70.90.2-0.70.10-0.3-0.8

Results are from Model: LW-9022th; RH Condition: 95%

 

Specifications

 

 

LW-9022-RT NATURAL CONVECTION AND CONSTANT TEMPERATURE CHAMBER

    1. Case Material: Clear Acrylic, 20 mm in Thickness
    2. Temperature Range: Ambient +3°C ~ 70°C
    3. Temperature Accuracy: ±5°C
    4. Temperature Uniformity: ±2°C
    5. Temperature Controller: PID
    6. Height-Adjust Platform: 50*100 cm (motor-driven)
    7. Power Source: AC220V, 15A, 3 Phases
    8. Chamber: 146 (W) * 116 (D) *176 (H) cm
    9. Door 60 * 100 cm * 2

 

 

LongWin offers a variety of models for demonstrations and tests. If you need accommodations that you do not see here on our website, please email us and we will try our best to provide you with the best instrument and tests.